Cardiac
Main Page
|
Specific
anatomic region
|
Cardiac
CTA |
|
Application
|
Test
bolus |
|
Author
|
Elliot
K Fishman, MD
|
|
Reference
Source
|
JHU
Protocol
|
|
Scanner
Used
|
Siemens
Sensation 64
|
|
KV/Effective
mAs/ Rotation time (sec)
|
120/
40 / .5 |
|
Detector
Collimation (mm)
|
10
mm |
|
Slice
thickness (mm)
|
10
mm |
|
Pitch
|
N/A
- 0 table feed |
|
Kernel
|
B20 |
|
Reconstruction
interval
|
10mm
|
|
Image
order (acquisition)
|
N/A
|
|
Oral
contrast
|
N/A
|
|
IV
contrast volume and type
|
Visipaque
320 20cc / 40cc Saline flush |
|
Injection
rate
|
4cc/sec |
|
Scan
delay (sec)
|
10
sec |
|
3D
technique used
|
N/A
|
|
Comment:
Uses 2 sec cycle time. Test bolus used to determine patient
specific delay time for CTA. Delay time + 6 sec is ideal.
|
|
|